Parameters
Comments
Usage
[SENSe:]STRain:EXCitation?
Parameters
Comments
Usage
Chapter 7
Parameter
Parameter
Name
Type
excite_v
numeric (float32)
ch_list
channel list (string)
•
The <ch_list> parameter must specify the channel used to sense the bridge
voltage, not the channel position on a Bridge Completion SCP.
•
Related Commands: SENSE:STRAIN:..., SENSE:FUNC:STRAIN...,
MEAS:VOLT:EXCitation
•
*RST Condition: 3.9 V
•
Send with VXIplug&play Function: hpe1422_cmd(...)
STRAIN:EXC Meas_excV,(@107)
STRAIN:EXC Meas_excV,(@10022)
[SENSe:]STRain:EXCitation? (@<channel>)
currently set for the sense channel specified by <channel>.
Parameter
Parameter
Name
Type
channel
channel list (string)
•
The <channel> parameter must specify a single channel only.
•
Returned Value: Numeric value of excitation voltage. The data type is float32.
•
Related Commands: SENS:STRAIN:EXCitation, MEAS:VOLT:EXCitation
•
Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)
STRAIN:EXC? (@107)
enter statement here
Range of
Values
0.01 - 99
100 - 15731
set measured excitation voltage for
channel 7
set excitation voltage for VT1529A/B
channel 0022
returns the excitation voltage value
Range of
Values
100 - 15731
query excitation voltage for channel 7
returns the excitation voltage set by
STR:EXC
VT1422A Command Reference
[SENSe]
Default
Units
volts
none
Default
Units
none
365