DIAGnostic:TEST:REMote:NUMber?

Parameters
Comments
Chapter 7
Returned Value: DIAG:REM:USER:DATA? returns an IEEE definite length
data block which represents an array of 894, int16 values.
RST Condition: Stored values not changed by *RST
Send with VXIplug&play Function: hpe1422_cmdInt16Arr_Q(...)
DIAGnostic:TEST:REMote:NUMber? <test_num>,<iterations>,(@<channel>)
executes a selected self-test number on a single Remote Signal Conditioning Unit
connected through the VT1539A SCP. See DIAG:TEST:REM:SELF? for details of
each test.
Parameter
Parameter
Name
Type
test_num
numeric
iterations
numeric
channel
channel list (string)
The <test_num> parameter specifies the test to perform. For explanations of the
test numbers, See "DIAGnostic:TEST:REMote:SELFtest?" on page 288.
The <iterations> parameter specifies the number of times to perform a test.
The <channel> parameter may contain only any single channel number on a
Remote Signal Conditioning Unit. All channels on that RSCU will be tested.
Returned Value:
Value
0
1
-1
The data type for this returned value is int16.
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
Range of
1 - 32767
10000 - 15731
Meaning
test passed
number of times test
Error information in FIFO buffer.
failed
Couldn't start
Query the Error Queue (SYST:ERR?)
remote self-test
See error messages starting on page 453.
VT1422A Command Reference
DIAGnostic
Default
Values
1 - 5
Further Action
None
See error codes below.
Units
none