[Sense:]Strain:gfactor; [Sense:]Strain:gfactor <Gage_Factor>,(@<Ch_List>); [Sense:]Strain:gfactor? (@<Channel>) - VXI VT1422A User Manual

Remote channel multi-function dac module user’s and scpi programming manual
Table of Contents

Advertisement

[SENSe:]STRain:GFACtor

Parameters
Comments
Usage
[SENSe:]STRain:GFACtor?
Parameters
Comments
Usage
Chapter 7
[SENSe:]STRain:GFACtor <gage_factor>,(@<ch_list>)
to be used to convert strain bridge readings for the channels specified by <ch_list>.
Parameter
Parameter
Name
Type
gage_factor
numeric (float32)
ch_list
channel list (string)
The <ch_list> parameter must specify the SCP or RSC channel used to sense
the bridge voltage, not the channel position on a Bridge Completion SCP.
Related Commands: SENSE:STRAIN:GFAC?, SENSE:FUNC:STRAIN...
*RST Condition: Gage factor is 2
Send with VXIplug&play Function: hpe1422_cmd(...)
SENS:STRAIN:GFAC 3,(@100:107)
SENS:STRAIN:GFAC 2.2,(@10000:10931)
[SENSe:]STRain:GFACtor? (@<channel>)
the sense channel specified by <channel>.
Parameter
Parameter
Name
Type
channel
channel list (string)
Returned Value: Numeric value of gage factor. The data type is float32.
The <channel> parameter must specify a single channel only.
Related Commands: STRAIN:GFACTOR
Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)
STRAIN:GFAC? (@107)
enter statement here
specifies the gage factor
Range of
Values
1 - 5
100 - 15731
set gage factor for channels 0 through 7
set gage factor for VT1529A/B channels
0000 through 0931 (128 channels)
returns the gage factor currently set for
Range of
Values
100 - 15731
query gage factor for channel 7
returns the gage factor set by STR:GFAC
VT1422A Command Reference
[SENSe]
Default
Units
none
none
Default
Units
none
367