SOURce:PULSe:WIDTh?

Parameters
Comments

SOURce:VOLTage[:AMPLitude]

Parameters
Comments
Chapter 7
The variable frequency control for this channel is provided by the algorithm
language. When the algorithm executes an assignment statement to this channel,
the value assigned will be the frequency setting. For example:
O143 = 2000 /* set channel 43 to 2 kHz */
Send with VXIplug&play Function: hpe1422_cmd(...)
SOUR:PULS:WIDTH 2.50E-3,(@143)
SOURce:PULSe:WIDTh? (@<ch_list>)
frequency modulated pulse channel.
Parameter
Parameter
Name
Type
channel
string
The <channel> parameter must specify a single channel.
If the channels specified are not on a Frequency/Totalize SCP, an error will
be generated.
Returned Value: returns the numeric pulse width. The type is float32.
Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)
SOURce:VOLTage[:AMPLitude] <-offset_v>,(@<ch_list>)
reduce bridge offset voltage present at the dynamic strain "Buffered Output"
channel connectors.
Parameter
Parameter
Name
Type
-offset_v
numeric (float32)
ch_list
string
If the channels specified are not on a VT1529A/B, an error will be generated.
*RST Condition: SOUR:VOLT 0,(@ <all VT1529A/B channels>)
set fixed pulse width of 2.5 ms on
channel 43
returns the fixed pulse width value on a
Range of
Values
100 - 163
can be used to
Range of
Values
-1.651 to +1.664 | MIN | MAX
(resolution 13 mV)
10000 - 15731
VT1422A Command Reference
SOURce
Default
Units
none
Default
Units
volts
none
377