Source:pulm:state; Source:pulse:period; Source:pulm[:State]? (@<Channel>); Source:pulse:period <Period>,(@<Ch_List>) - VXI VT1422A User Manual

Remote channel multi-function dac module user’s and scpi programming manual
Table of Contents

Advertisement

Comments

SOURce:PULM:STATe?

Parameters
Comments

SOURce:PULSe:PERiod

Parameters
Comments
Chapter 7
This command is coupled with the SOURce:FM command. If the FM state is
enabled then the PULM state is disabled. If the PULM state is enabled then the
FM state is disabled. If both the FM and the PULM states are disabled then the
PULSe channel is in the single pulse mode.
If the channels specified are not on a Frequency/Totalize SCP, an error will
be generated.
*RST Condition: SOUR:PULM:STATE OFF
Send with VXIplug&play Function: hpe1422_cmd(...)
SOURce:PULM[:STATe]? (@<channel>)
state for the PULSe channel in <channel>.
Parameter
Parameter
Name
Type
channel
string
The <channel> parameter must specify a single channel.
Returned Value: returns 1 (on) or 0 (off). The type is int16.
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
SOURce:PULSe:PERiod <period>,(@<ch_list>)
on a pulse width modulated pulse channel. This sets the frequency (1/period) of the
pulse-width-modulated pulse train.
Parameter
Parameter
Name
Type
period
numeric (float32)
ch_list
string
If the channels specified are not on a Frequency/Totalize SCP, an error will
be generated.
*RST Condition: SOUR:FM:STATE OFF and SOUR:PULM:STATE OFF
Related Commands: SOUR:PULM:STATE, SOUR:PULS:POLarity
returns the pulse width modulated mode
Range of
Values
100 - 163
sets the fixed pulse period value
Range of
Values
25E-6 to 7.8125E-3
(resolution 0.238 μs)
100 - 163
VT1422A Command Reference
SOURce
Default
Units
none
Default
Units
seconds
none
375

Advertisement

Table of Contents
loading

Table of Contents