Diagnostic:calibration:setup[:Mode]; Diagnostic:calibration:setup[:Mode] <Mode - VXI VT1422A User Manual

Remote channel multi-function dac module user’s and scpi programming manual
Table of Contents

Advertisement

DIAGnostic:CALibration:SETup[:MODE]

Parameters
Comments
Usage

DIAGnostic:CALibration:SETup[:MODE]?

Comments
Chapter 7
DIAGnostic:CALibration:SETup[:MODE] <mode>
use for analog output SCPs like the VT1531A and VT1532A when *CAL? or
CAL:SET are executed.
Parameter
Parameter
Name
Type
mode
boolean (uint 16)
When <mode> is set to 1 (the *RST Default) channels are calibrated using the
Least Squares Fit method to provide the minimum error overall (over the entire
output range). When <mode> is 0, channels are calibrated to provide the
minimum error at their zero point. See the SCPs User's Manual for its accuracy
specifications using each mode.
Related Commands: *CAL?, CAL:SET, DIAG:CAL:SET:MODE?
*RST Condition: DIAG:CAL:SET:MODE 1
Send with VXIplug&play Function: hpe1422_cmd(...)
set analog DAC SCP cal mode for best zero accuracy
DIAG:CAL:SET:MODE 0
*CAL?
DIAGnostic:CALibration:SETup[:MODE]?
mode for analog output DAC SCPs.
Returns a 1 when channels are calibrated using the Least Squares Fit method to
provide the minimum error overall (over the entire output range). Returns a 0
when channels are calibrated to provide the minimum error at their zero point.
See the SCPs User's Manual for its accuracy specifications using each mode.
The data type is int16.
Related Commands: DIAG:CAL:SET:MOD, *CAL?, CAL:SET
*RST Condition: DIAG:CAL:SET:MODE 1
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
sets the type of calibration to
Range of
Values
0 | 1
set mode for best zero cal
start channel calibration
returns the currently set calibration
VT1422A Command Reference
DIAGnostic
Default
Units
volts
277