DIAGnostic:CALibration:SETup[:MODE]
Parameters
Comments
Usage
DIAGnostic:CALibration:SETup[:MODE]?
Comments
Chapter 7
DIAGnostic:CALibration:SETup[:MODE] <mode>
use for analog output SCPs like the VT1531A and VT1532A when *CAL? or
CAL:SET are executed.
Parameter
Parameter
Name
Type
mode
boolean (uint 16)
•
When <mode> is set to 1 (the *RST Default) channels are calibrated using the
Least Squares Fit method to provide the minimum error overall (over the entire
output range). When <mode> is 0, channels are calibrated to provide the
minimum error at their zero point. See the SCPs User's Manual for its accuracy
specifications using each mode.
•
Related Commands: *CAL?, CAL:SET, DIAG:CAL:SET:MODE?
•
*RST Condition: DIAG:CAL:SET:MODE 1
•
Send with VXIplug&play Function: hpe1422_cmd(...)
set analog DAC SCP cal mode for best zero accuracy
DIAG:CAL:SET:MODE 0
*CAL?
DIAGnostic:CALibration:SETup[:MODE]?
mode for analog output DAC SCPs.
•
Returns a 1 when channels are calibrated using the Least Squares Fit method to
provide the minimum error overall (over the entire output range). Returns a 0
when channels are calibrated to provide the minimum error at their zero point.
See the SCPs User's Manual for its accuracy specifications using each mode.
The data type is int16.
•
Related Commands: DIAG:CAL:SET:MOD, *CAL?, CAL:SET
•
*RST Condition: DIAG:CAL:SET:MODE 1
•
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
sets the type of calibration to
Range of
Values
0 | 1
set mode for best zero cal
start channel calibration
returns the currently set calibration
VT1422A Command Reference
DIAGnostic
Default
Units
volts
277