7. CLEANING AND DISINFECTING

7.1. HANDPIECE

We advise to clean and disinfect the handpiece
after each treatment
1. Clean and Disinfect with a cleaning and disinfectant
wipe such as
Surface Wipes Zero from
or equivalent (quaternary ammonium as active
molecule).

7.2. APPLICATORS

Ensure the applicator is cleaned and disinfected
prior to treating a patient. It should take only
about 5 minutes.
26
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 26
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 26
Helvemed,
Figure 53
1. Disassemble the applicator, the nose, the retaining
nut and the o'rings.
Figure 54
2. Clean and disinfect the applicator with a cleaning and
disinfectant wipe such as
Helvemed, or equivalent (quaternary ammonium as
active molecule).

7.3. CONSOLE

We advise to clean and disinfect the console at
the end of the day
1. Turn off the console.
2. Disconnect the power supply connector.
3. Clean and disinfect the surfaces using a cleaning and
disinfectant wipe such as
Helvemed, or equivalent (quaternary ammonium as
active molecule).
Surface Wipes Zero from
Surface Wipes Zero from