16. TECHNICAL DESCRIPTION

MANUFACTURER
MODEL
POWER SUPPLY
CLASSIFICATION EN 60601-1
RADIO COMMUNICATION
MDD 93/42 EEC CLASSIFICATION
OPERATING PRESSURE
OPERATING FREQUENCY
IP CLASSIFICATION IEC 60601-1
FUSE
WEIGHT
DIMENSION (HXWXD)
OPERATING CONDITIONS
STORAGE
TRANSPORT CONDITIONS
SERVICE LIFE
VIBRATION
38
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 38
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 38
EMS ELECTRO MEDICAL SYSTEMS SA, CH-1260 Nyon, Switzerland
DOLORCLAST
Radial Shock Waves
®
100-240 VAC, 50-60 Hz, 500VA
System: class I
Applied part type BF
RFID 13.56MHz, < 20dBuA/m
Console, handpiece: class IIb
1 - 4 bar
1 - 25 Hz
Console: IP21
Handpiece: IP20
T5A 250V High breaking capacity
Console: 14.5 kg
Handpiece: 0.85 kg
Cart: 18 kg
Console: 180mm x 340mm x 450mm
Handpiece: 113mm x 250mm x 51mm
Cart: 913mm x 460mm x 541mm
+10 °C to + 30 °C
Relative humidity: 35 % to 65 %
Max. altitude: 3000 m
Atmospheric pressure: 700 hPa to 1060 hPa
Temperature: +1°C to +30°C
Relative Humidity: max. 85%
Console and its accessories
Temperature: +1°C to +30°C
Relative Humidity: max. 85%
Handpiece
Temperature: -29°C to +38°C
Relative Humidity: max. 85%
Console: 7 years
Handpiece: 5 000 000 shots
3.55 m/s2
equivalent to about 120 treatments of 3000 impulses at 4bar/25Hz per
day and per operator