Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The DOLORCLAST
®
The customer or the user of the DOLORCLAST
EMISSIONS TEST
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonics emissions
IEC 61000-3-2
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The DOLORCLAST
below. The customer or the user of DOLORCLAST
IMMUNITY TEST
Electrostatic
± 8 kV contact
discharge (ESD)
± 2, ± 4, ± 8, ± 15 kV air
IEC 61000-4-2
Electrical fast
± 2 kV, 100 kHz
transient / burst
for power supply lines*
IEC 61000-4-4
± 1 kV, 100 kHz
for input/output lines*
Surge
±0.5, ±1 kV
IEC 61000-4-5
line(s) to line(s)*
±0.5, ±1, ± 2 kV
line(s) to earth*
Power frequency
30 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Voltage dips,
0 % UT: 0.5 cycle
short interruptions
At 0°, 45°, 90°, 135°,
and voltage
180°, 225°, 270° and
variations on
315°
power supply
0 % UT: 1 cycle
input lines
At 0°
IEC 61000-4-11
0 % UT: 250/300 cycles
At 0°
70 % UT: 25/30 cycles
At 0°
34
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 34
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 34
Radial Shock Waves is intended for use in the electromagnetic environment specified below.
®
COMPLIANCE
Group 1
The DOLORCLAST
function. Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Class B
The DOLORCLAST
ments, including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings used for
Class B
domestic purposes.
Complies
Radial Shock Waves is intended for use in the electromagnetic environment specified
®
IEC 60601
COMPLIANCE LEVEL
TEST LEVEL
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
± 2 kV, 100 kHz
for power supply lines*
± 1 kV, 100 kHz
for input/output lines*
±0.5, ±1 kV
line(s) to line(s)*
±0.5, ±1, ± 2 kV
line(s) to earth*
30 A/m
0 % UT: 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT: 1 cycle
At 0°
0 % UT: 250/300 cycles
At 0°
70 % UT: 25/30 cycles
At 0°
Radial Shock Waves should assure that it is used in such an
environment.
ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
Radial Shock Waves uses RF energy only for its internal
®
Radial Shock Waves is suitable for use in all establish-
®
Table 3
Radial Shock Waves should ensure that it is used in such an
®
environment.
ELECTROMAGNETIC ENVIRONMENT -
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Mains power quality should be that of typical
commercial or hospital environment.
*Not applicable for DC and I/O if cable < 3 m
Mains power quality should be that of typical
commercial or hospital environment.
*Not applicable for DC and I/O if cable < 3 m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the DOLORCLAST
requires continued operation during power
mains interruptions, it is recommended that the
SWISS DOLORCLAST
powered from an uninterruptible power supply or
a battery.
*Not applicable for DC and I/O if cable < 3 m
UT is the AC mains voltage (100-240) prior to
application of the test level.
Table 4
GUIDANCE
Radial Shock Waves
®
Radial Shock Waves be
®