EMS DOLORCLAST Instructions For Use Manual page 28

Radial shock waves
Table of Contents

Advertisement

Available languages

Available languages

  • ENGLISH, page 3
e. Pull on the piston to remove it from the handpiece.
3. Discard the worn piston according to the regulations
that apply in your country.
4. Unpack the new piston and make sure to remove
the piston cap before installing the new piston on the
handpiece.
28
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 28
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 28
Figure 59
Figure 60
Figure 61
5. Install the new piston on the handpiece.
6. Re-assemble the handpiece.
8.1.2. Applicator O-ring
EMS recommends to change the o'rings when
the piston is changed
8.1.2.1. For 5-mm, 10-mm and 15-mm Applicators
1. Disassemble the applicator.
Figure 62
2. If the O-ring is damaged or worn, it must be replaced
immediately.
Figure 63
3. Use the O-ring maintenance kit and select the relevant
O-ring according to the applicator size.
25/03/2021 17:07