Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The DOLORCLAST
®
The customer or the user of the DOLORCLAST
Portable and mobile RF communications equipment should be used no closer to any part of the DOLORCLAST
Waves, including cables, than the recommended separation distance calculated from the equation applicable to the frequency
of the transmitter.
These guidelines may not apply to all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects and people. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radios, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered.
IMMUNITY
TEST
TEST LEVEL
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
outside ISM bands
and radio amateur
band *
6 Vrms
150 kHz to 80 MHz
in ISM bands and
radio amateur
band*
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
80 % AM at 1KHz
Proximity
27 V/m
fields from
380-390 MHz
RF wireless
50 % PM 18 Hz
commu-
nication
28 V/m
equipment
430-470 MHz
IEC 61000-4-3
FM ±5 kHz
deviation,
1 kHz sine
9 V/m
704-787 MHz
50 % PM 217 Hz
28 V/m
800-960 MHz
50 % PM 18 Hz
28 V/m
1700-1990 MHz
50 % PM 217 Hz
28 V/m
2400-2570 MHz
50 % PM 217 Hz
9 V/m
5100-5800 MHz
50 % PM 217 Hz
*The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 - 6.795 MHz, 13.553 - 13.567 MHz, 26.957 -
27.283 MHz and 40.66 - 40.7 MHz. The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz - 2 MHz, 3.5 - 4.0 MHz, 5.3 -
5.4 MHz, 7 - 7.3 MHz, 10.1 - 10.15 MHz, 14 - 14.2 MHz, 18.07 - 18.17 MHz, 21.0 - 21.4 MHz, 24.89 - 24.99 MHz, 28.0 - 29.7 MHz and 50.0 -
54.0 MHz.
If the measured field strength in the location in which the DOLORCLAST
level above, the DOLORCLAST
additional measures may be necessary, such as reorienting or relocating the DOLORCLAST
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 35
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 35
Radial Shock Waves is intended for use in the electromagnetic environment specified below.
IEC 60601
COMPLIANCE
LEVEL
6 Vrms
150 kHz to 80 MHz
outside ISM bands
and radio amateur
band *
6 Vrms
150 kHz to 80 MHz
in ISM bands and
radio amateur
band*
3 V/m
80 MHz to 2.7 GHz
80 % AM at 1KHz
27 V/m
380-390 MHz
50 % PM 18 Hz
28 V/m
430-470 MHz
FM ±5 kHz
deviation,
1 kHz sine
9 V/m
704-787 MHz
50 % PM 217 Hz
28 V/m
800-960 MHz
50 % PM 18 Hz
28 V/m
1700-1990 MHz
50 % PM 217 Hz
28 V/m
2400-2570 MHz
50 % PM 217 Hz
9 V/m
5100-5800 MHz
50 % PM 217 Hz
Radial Shock Waves should be observed to verify normal operation. If abnormal performance is observed,
®
Radial Shock Waves should ensure that it is used in such an
®
environment.
ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
If the measured field strength in the location in which the
DOLORCLAST
Radial Shock Waves is used exceeds the
®
applicable RF compliance level above, the SWISS DOLOR-
CLAST
Radial Shock Waves should be observed to verify
®
normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting
or relocating the DOLORCLAST
Minimum separation distance shall be calculated by following
equation:
E is the immunity test level in [V/m]
d is the minimum separation in [m]
P is the maximum power in [W]
RF wireless equipment maximum output power and
separation distance tested (at 30 cm):
TETRA 400: max 1.8 W
GMRS 460, FRS 460: max 2 W
LTE Band 13, 17: max 0.2 W
GSM 800/900: max 2 W
TETRA 800: max 2 W
iDEN 820: max 2 W
CDMA 850: max 2 W
LTE Band 5: max 2 W
GSM 1800/1900: max 2 W
CDMA 1900: max 2W
DECT: max 2 W
LTE Band 1,3,4,25: max 2 W
UMTS: max 2 W
Bluetooth: max 2W
WLAN 802.11b/g/n: max 2 W
RFID 2450: max 2 W
LTE Band 7: max 2 W
WLAN 802.11 a/n: max 0.2 W
Interference may occur in the vicinity of equipment marked
with the following symbol:
Radial Shock Waves is used exceeds the applicable RF compliance
®
Radial Shock Waves
®
Table 5
Radial Shock
®
Radial Shock Waves.
®
E = 6 P
d
35