Burst Modes - EMS DOLORCLAST Instructions For Use Manual

Radial shock waves
Table of Contents

Advertisement

Available languages

Available languages

  • ENGLISH, page 3
1. Before starting treatment or after a pause, press the
Ramp-Up button twice to enable Ramp-Up mode 2.
The LED signifies that the mode is enabled.
1.5
bar
25
Hz
3000
Impulses
SMART PROTOCOL
2. Short-press the trigger on the handpiece to start
treatment.
3. The pressure on the console will automatically
increase by 0.1 bar every 150 impulses until the
pressure reaches 4 bar.
4. The treatment will stop once a pressure of 4 bar is
reached, or once the remaining number of impulses
is reached.
If the patient begins to experience discomfort,
stop and adjust the treatment settings accor-
dingly.
To directly reach 4 bar, use the + button to in-
crease the pressure setting manually.
24
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 24
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 24
Figure 47
1.5
bar
25
Hz
3000
Impulses
SMART PROTOCOL
Figure 48

6.3. BURST MODES

Use the Burst modes to prevent the patient from devel-
oping tolerance to certain mechanical stimuli. The Burst
modes are specifically designed to alternate a preset
frequency and a maximum frequency during treatment.
6.3.1. Burst Mode 1
Using Burst mode 1 during treatment will automatically
alternate between 4 seconds at the set frequency and
2 seconds at 25 Hz. The pressure setting will remain
constant.
Figure 49
1. Press the Burst button to enable Burst mode 1. The
LED signifies that the mode is enabled.
1.5
bar
1.5
bar
25
25
Hz
Hz
1000
Impulses
1000
Impulses
SMART PROTOCOL
SMART PROTOCOL
Figure 50
2. To activate the treatment using Burst mode 1, press
the trigger on the handpiece.
3. Burst mode will stop automatically when all treatment
impulses are delivered.
25/03/2021 17:07
25/03/2021 17:07