Recommended separation distances between portable and mobile RF communications
The DOLORCLAST
radiated RF disturbances are controlled. The customer or the user of the DOLORCLAST
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the DOLORCLAST
below, according to the maximum output power of the communications equipment.
Rated
150 kHz to 80 MHz outside
maximum
ISM and radio amateur bands*
output power
of transmitter
0.01 W
0.1 W
1 W
10 W
100 W
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters [m] can
be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts [W] according to the transmitter manufacturer.
*The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 - 6.795 MHz, 13.553 - 13.567 MHz,
26.957 - 27.283 MHz and 40.66 - 40.7 MHz. The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz - 2 MHz, 3.5 -
4.0 MHz, 5.3 - 5.4 MHz, 7 - 7.3 MHz, 10.1 - 10.15 MHz, 14 - 14.2 MHz, 18.07 - 18.17 MHz, 21.0 - 21.4 MHz, 24.89 - 24.99 MHz,
28.0 - 29.7 MHz and 50.0 - 54.0 MHz.
The use of accessories and cables other than those specified or sold by EMS as replacement parts may result
in increased emissions or decreased immunity of this product.
CABLES AND
ACCESSORIES
Handpiece
Sector cord
The DOLORCLAST
Radial Shock Waves has neither life sustaining functions nor diagnostic of life supporting functions.
®
The following functions are observed:
Continuous work of pneumatic shocks,
No changes of mode,
No changes in set-values.
36
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 36
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 36
equipment and the DOLORCLAST
Radial Shock Waves is intended for use in an electromagnetic environment in which
®
SEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER [M]
150 kHz to 80 MHz outside
ISM and radio amateur bands*
d = 1.0√P
0.1
0.32
1.0
3.2
10
Compliant Cables and Accessories
MAXIMUM
LENGTH
< 2.9 m
CISPR 11 Class B / Group 1: RF electromagnetic disturbance
> 2.9 m
IEC 61000-4-2 Electrostatic discharge (ESD)
IEC 61000-4-3 Electromagnetic fields radiated by radio-frequencies
IEC 61000-4-4 Electric fast transient / burst
IEC 61000-4-5 Surge
IEC 61000-4-6 Disturbances induced by radio-frequency fields
IEC 61000-4-8 Power frequency magnetic field (50/60 Hz)
IEC 61000-4-11 Voltage dips, short interruptions and voltage variations
Essential Performance
Radial Shock Waves
®
Radial Shock Waves as recommended
®
80 MHz to 2700 MHz (for define RF
Wireless transmitters see table before)
d = 1.0√P
0.1
0.32
1.0
3.2
10
E = 6 P
d
Table 6
COMPLIES WITH
Table 7
Radial Shock Waves
®
d = 0.60√P
0.06
0.19
0.60
1.9
6.0