3.3. SOFTWARE AND HANDPIECE
INFORMATION (NUMBERS OF USE)
1. Press and hold the three
and hold them for 5 seconds.
2. The following screen appears. The Software version
is displayed on page 1.
Main (ES-106):
Console HMI (ES-107):
Handpiece (ES-1002):
3. Handpiece information is displayed on page 2. Press
to display page 2.
Piston: 0 treatments, 0 impulses
(1000000 max impulses)
Means: 0.0 bar, 0 Hz, 0 impulses
Handpiece: 0 treatments, 0 impulses
Means: 0.0 bars, 0 Hz, 0 impulses
EMS does not guarantee the performance of the
treatment if the piston has reached its maximum
recommended impulses.
4. To exit the Software and Handpiece information
screen, press and hold the three
taneously and hold them for 5 seconds.
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 13
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 13
buttons simultaneously
Software version
1.1.3
0.8.18
0.2.2
Page 1
Figure 15
Handpiece information
Page 2
Figure 16
buttons simul-

3.4. SETTING THE LANGUAGE

1. Press the three special mode buttons simultaneously
and hold them for 5 seconds.
2.0
bar
15
Hz
2500
Impulses
SMART PROTOCOL
Figure 17
2. The following screen will appear:
Do you want to perform a factory reset?
No
Figure 18
Changing the language will reset to factory set-
tings. If you have saved a USER protocol, it will
be deleted.
3. Press
to confirm that you wish to perform a
factory reset/change the language.
Yes