Ramp-Up Modes - EMS DOLORCLAST Instructions For Use Manual

Radial shock waves
Table of Contents

Advertisement

Available languages

Available languages

  • ENGLISH, page 3
ANALGESIC
1000
SMART PROTOCOL
3. To start treatment with Analgesic mode 2, short-press
the trigger on the handpiece.
Press the trigger to pause the Analgesic mode.
4. After 1000 impulses, the device stops and returns to
preset settings to continue treatment.
5. Press the trigger on the handpiece to start treatment.

6.2. RAMP-UP MODES

The Ramp-Up modes a re designed to help you treat a
patient at maximum pressure without having to manually
adjust the pressure. Shockwave treatments are dose
dependent. The best treatment outcomes will be reached
at the maximum possible pressure tolerated by the
patient.
If treatment is paused, the Ramp-Up mode is disabled.
Press the trigger again to continue treatment using the
last pressure threshold reached.
6.2.1. Ramp-Up Mode 1
Ramp-Up mode 1 will automatically increase the pressure
by 0.1 bar every 150 impulses until a total increase of 0.5
bar is reached, or until the remaining number of impulses
is reached.
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 23
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 23
[0 mJ/mm
]
2
25
ANALGESIC
[0 mJ/mm
2
]
Hz
25
Hz
1000
Impulses
Impulses
SMART PROTOCOL
Figure 44
Figure 45
1. Before starting treatment or after a pause, press the
Ramp-Up button to enable Ramp-Up mode 1. The
LED signifies that the mode is enabled.
1.5
bar
1.5
25
Hz
750
750
Impulses
SMART PROTOCOL
SMART PROTOCOL
Figure 46
2. Short-press the trigger on the handpiece to start
treatment.
3. After the pressure has increased by a total of 0.5 bar
or the remaining number of impulses is reached, the
Ramp-Up mode is completed.
6.2.2. Ramp-Up Mode 2
Ramp-Up mode 2 will automatically increase the pressure
by 0.1 bar every 150 impulses until a total increase
of 4 bar is reached, or until the specified number of
impulses is reached. After this, the system will display
the previous pressure setting.
bar
25
Hz
Impulses
23
25/03/2021 17:07
25/03/2021 17:07

Advertisement

Table of Contents
loading