1. Ensure the O-ring is still in place with the spare appli-
cator. If the O-ring is disconnected from the applicator,
replace it.
If the O-ring is incorrectly positioned, this will
cause the handpiece to malfunction.
2. Loosen and remove the applicator from the handpiece.
3. Tighten the new applicator onto the handpiece.
Ensure the applicator is completely tightened.
4. The handpiece is ready to use.
16
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 16
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 16
Figure 24
Figure 25
Figure 26
4.2.2. Adjusting the Pressure
To adjust the shock wave depth and the energy trans-
mitted into the tissue, increase or decrease the pressure.
The pressure can be selected between 1 and 4 bar, in
increments of 0.1 bar.
3.0
bar
15
Hz
2000
Impulses
SMART PROTOCOL
Figure 27
1. To increase the pressure, press the
2. To decrease the pressure, press the
3. Press and hold the
button to obtain an increment
of 0.2 bar.
It is possible to adjust the pressure during treat-
ment.
4.2.3. Adjusting the Frequency
To adjust the number of impulses per second, increase or
decrease the frequency. The frequency can be selected
between 1 and 25 Hz, in increments of 1 Hz.
3.0
bar
15
Hz
2000
Impulses
SMART PROTOCOL
Figure 28
1. To increase the frequency, press the
2. To decrease the frequency, press the
It is possible to adjust the frequency during treat-
ment.
button.
button.
button.
button.