8. MAINTENANCE

Should legal provisions in your country specify
maintenance intervals, these must be observed.
The console and handpiece may need to be re-
turned for periodic servicing

8.1. HANDPIECE

8.1.1. Piston
When the piston has reached 1'000'000 shots, the following
screens appears, indicating that periodic replacement has
to be done on the piston using the BLUE maintenance kit.
E008 - The piston has reached its
maximum recommended usage
Replace the piston using the maintenance
kit for the BLUE handpiece:
Press any button to continue
The QR code that will appear on your console
screen will lead to an EMS website allowing you
to order a new piston.
To change the piston, proceed as following:
1. Disconnect the handpiece from the device
2. Disassemble the piston
a. Unscrew the applicator and remove it from the
handpiece.
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 27
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 27
2.0
bar
15
Hz
2500
Impulses
SMART PROTOCOL
Figure 55
b. Unscrew the retaining nut and remove it from the
handpiece.
Figure 56
c. Pull on the protective cover to remove it from the
handpiece.
Figure 57
d. Push the end of the bayonet towards the handpiece
and turn it to release the bayonet.
Figure 58
27
25/03/2021 17:07
25/03/2021 17:07