Programming Infusions
Note:
Note:

Reviewing PCA Profile Parameters

07-19-12-353
If you enter or accept a dosage of 0.0, the Clinician
Dose function is canceled. If you make an error,
press NO to change your entry.
When you press YES to accept an entered dosage
parameter, the pump displays the Clinician Dose
confirmation screen:
CLINICIAN
Dose:
alternating with:
***
Press RUN/HOLD to begin the Clinician Dose.
4.
As the clinician dose is being administered, the dis-
played dosage decreases. When the bolus dose vol-
ume reaches zero, the pump continues the infusion
at the current basal rate.
You have several options for reviewing PCA profile
information:
Review profile parameters, review PCA
(bolus) data, clear volume infused display
and review volume remaining while an
infusion is on hold.
Review PCA (bolus) information during an
infusion.
Review PCA (bolus) information after an
infusion has completed.