Delivery Options and Advanced Functions

Infusion Data Parameters in Programmable Mode

Note:
Note:
07-19-12-353
In Programmable Mode, access the Infusion Data
parameters by pressing
*HOLD* screen, the YES to PROGRAM screen or any
READY (Profile) screen.
See Table 9-1 for a list of the parameter limits.
1.
From a
*HOLD*
PROGRAM
The pump displays:
SET AUDIO VOLUME?
Press YES to review or change Audio Volume
2.
setting or NO to continue to Pressure. The pump
displays:
SET AUDIO VOLUME
(1 TO 9)
3.
Enter a number from 1 to 9 (9 is loudest). The pump
sounds a tone at the chosen volume. When you hear
the desired volume level, press YES to accept.
The pump displays:
SET PRESSURE?
Press YES to review or change the pressure setting,
4.
or press NO to continue to KO rate.
Options for downstream occlusion detection pressure
are
(20 psi ±5 psi) or
HI
Press YES to accept the current pressure setting, or
5.
press NO to switch to the other option, then YES to
accept.
The pump displays the KO rate screen:
6060™ Multi–Therapy Pump Operator's Manual
INFUSION DATA
,
READY (Profile
screen, press the INFUSION DATA key.
(8 psi ±5 psi).
LOW
from the
) or
YES to
5
5-21