Programming Continuous Profile

4-2
Power the pump on by pressing ON/OFF PUMP .
1.
The pump executes its self-test, then displays:
YES to Program
RUN to Repeat
Press YES . The pump displays the current
2.
downstream occlusion pressure setting, then
displays:
CONTINUOUS?
Press YES to select Continuous profile.
3.
DELAY DELIVERY?
Do one of the following:
4.
If you want to delay the start of the infusion,
press YES , enter the delay time in hours:minutes,
then press YES again.
If you want to start the infusion immediately after
programming the pump, press NO.
The pump displays:
PROGRAM in ml's?
Select the programming units as follows:
5.
To program in ml, press YES and go to step 6.
5.1
If not, press NO . The pump displays:
PROGRAM in mg's?
6060™ Multi–Therapy Pump Operator's Manual
07-19-12-353