Preventive Maintenance

Preventive Maintenance

Table 7-3
Preventive Maintenance
Check
Perform as required but recommended after every use.
Housing
Labels
Optional
PCA cord
Rear housing

Storing the Pump

7-12
Table 7-3 contains a schedule of basic maintenance
tasks that should be performed on the pump. If the
pump cannot be cleaned using the methods described
earlier or components are missing or damaged,
discontinue use and notify the appropriate authorized
service personnel. See "Service Information," 10-2.
Clean housing and front panel as recommended in the clean-
ing instructions in this section. Check for cracks or chips.
Clean as recommended in the cleaning instructions. Check
for scratches, cuts or obliterated words.
Verify that the cord is undamaged over its entire length and
that the PCA button is intact and functional.
Verify that there are no loose or missing parts and that con-
nectors and accessories are undamaged.
1.
To prevent exposure to dust and moisture, store the
pump in its shipping carton in a clean, dry
environment with air quality equivalent to or better
than normal warehouse air.
Store at -28.8 ° C to 60 ° C (-20 ° F to 140 ° F),
2.
10% to 95% humidity (non-condensing),
50 kPa to 106 kPa.
Remove the internal 9-volt batteries before storing
3.
the pump for long periods.
6060™ Multi–Therapy Pump Operator's Manual