Making the Pump Profile-Specific

Note:

Making the Pump Profile-Specific

5-10
For example, if the programmed concentration is 10
mg/ml and the infusion rate is 10 mg/hr, and you
change the concentration to 20 mg/ml, the pump will
perform the necessary calculations and increase the
rate of infusion to 20 mg/hr.
In some instances you may want to change the
concentration while keeping the infusion rate at its
originally programmed value.
In order to change the concentration, the pump must
be in Programmable Mode and at the
YES TO PROGRAM
concentration after an infusion has started running.
To change the concentration setting without
automatically recalculating the infusion rate:
At the
YES to PROGRAM
1.
2
6
The pump displays:
Set Global
Press YES . The pump displays:
2.
Maintain Rate
w/ Conc. Change?
Press YES .
3.
The Profile-Specific function lets you restrict the
profiles available for programming and delivery,
even in Programmable Mode. Selecting
profile makes that profile unavailable until the code
is reentered and the profile allowed. The choices are
listed in Table 5-1.
6060™ Multi–Therapy Pump Operator's Manual
screen. You cannot change the
6 4
Features?
screen, enter the code:
NO
for any
07-19-12-353