Programming Infusions
Changing Basal Rate and/or Bolus Parameters Without
Losing PCA Data
07-19-12-353
9.
The data is cleared. If the pump is on hold, it displays
the volume remaining in the infusion, then returns to
the
***HOLD***
continue the infusion.
If the infusion is running, the pump displays:
10.
PCA Level X
X.X
11.
If the infusion has completed, the pump displays:
LOCKOUT
RUN to Repeat
In PCA Security Levels 2 and 3, programming
changes to basal rate or bolus parameters during an
infusion normally reset the PCA data. To change
these parameters without losing accrued PCA data
and volumes, use the following procedure:
While the infusion is running, enter the code:
1.
6
1
2.
The pump displays:
Change Profile
Parameters?
Press YES . The pump displays:
3.
Rate
6060™ Multi–Therapy Pump Operator's Manual
screen. Press RUN/HOLD to
O
X.X
O
1
6
Basal
XX.X mg/hr
4
4-57