Using the Rechargeable External Battery Pack

Caution
07-19-12-353
The Rechargeable External Battery Pack contains
rechargeable power cells, and operates for about
twice as long as the internal 9-volt batteries. When
the Rechargeable External Battery Pack is used to
power the pump, the internal 9-volt batteries are not
used, saving their power for later use.
Maintain Rechargeable External Battery
Pack charge by keeping the Rechargeable
External Battery Pack plugged into the
Battery Eliminator/Charger when not in use.
Always keep two 9-volt batteries installed in
the pump.
Store Rechargeable External Battery Pack in
a cool place.
Disconnect the Battery Eliminator/Charger
from the Rechargeable External Battery Pack
before unplugging the Battery
Eliminator/Charger from the wall.
Recharge immediately after use. If the
Rechargeable External Battery Pack is
repeatedly allowed to drain completely, its
life will be shortened.
The fast charge light on the Battery
Eliminator/Charger indicates that it is in the
high state charge mode. When the fast charge
light goes off it does not indicate that the
Rechargeable External Battery Pack is fully
charged. Full charge time is 24 hours.
Charging for 24 hours maximizes battery
capacity and lifetime.
Never expose a discharged battery to freezing
temperatures.
6060™ Multi–Therapy Pump Operator's Manual
7
7-5