Programming Continuous Profile

Titration in Continuous Profile

4-6
To deliver the programmed infusion at a later time,
11.
press the ON/OFF PUMP key to turn the pump off.
When you want to begin the programmed infusion,
follow the directions under "Shortcut Delivery of a
Programmed Infusion," 5-2.
To begin the infusion immediately, press YES. The
pump displays the Final Confirmation screen:
CONTINUOUS
alternating with:
RATE/hr (ug) AMT
XXX
12.
Review the settings shown in the Final Confirmation
screen and do one of the following:
Press NO/REVIEW to review parameters again.
Press RUN to begin the infusion. The green
indicator light blinks and a flashing arrow is
displayed on the screen to show that the pump is
delivering fluid.
Titration mode allows you to change the delivery rate
and/or the volume of a continuous infusion during
delivery or before the start of an infusion.
There are three ways to titrate a continuous infusion:
Change rate without stopping the infusion.
Change rate and/or volume when the
infusion is put on hold.
Change rate and/or volume from the Final
Confirmation screen before the infusion
begins.
6060™ Multi–Therapy Pump Operator's Manual
Press Run