Using The Pump To Prime - baxter 6060 2M9832 Operator's Manual

Multi–therapy ambulatory infusion pump
Table of Contents

Advertisement

Using Administration Sets

Using the Pump to Prime

! WARNING !
Note:
3-4
The administration set MUST NOT be connected to
the patient while priming.
Use aseptic technique when priming and loading the
administration set.
To prime the administration set using the pump:
1.
Load the administration set as described in "Loading
the Administration Set" on page 3-3.
Turn the pump on. Program the desired infusion (see
2.
the profile programming sections in Chapter 4) and
advance the pump to the
screen. See "Ready Screens / Final Confirmation
Screens" on page 3-16.
If the pump is in Lockout Mode, see "Programmable
Mode / Lockout Mode," 5-16 for instructions on
starting an infusion in Lockout Mode.
At the Final Confirmation screen, press
3.
PRIME/BOLUS .
The pump displays one of the following:
Prime Set?
or:
Prime Set?
enter code
Press YES or enter the following code if the pump is
4.
in lockout:
9
1
6060™ Multi–Therapy Pump Operator's Manual
Final Confirmation