Clearing The Memory; Changing Concentration Without Affecting Programmed Rate - baxter 6060 2M9832 Operator's Manual

Multi–therapy ambulatory infusion pump
Table of Contents

Advertisement

Delivery Options and Advanced Functions

Clearing the Memory

Changing Concentration Without Affecting
Programmed Rate
07-19-12-353
All programmed values stay in the pump's memory
until they are changed (reprogrammed) or cleared
using the following method. Clearing the memory
erases all programmed delivery values and profile
parameters. Only the infusion data parameters (audio
volume, KO rate, pressure setting, display language,
air detection setting and up occlusion) are not
cleared.
At the
YES to PROGRAM
1.
6
0
The pump displays:
MEMORY CLEARED
then displays:
YES to Program
RUN to Repeat
To confirm that the memory has been cleared, press
2.
RUN/HOLD (the shortcut to begin a preprogrammed
infusion).
The pump should display
there are no programmed profiles available for
delivery.
When programming the pump in mg or µ g, you must
enter a concentration so the pump can calculate the
equivalent in ml. If the concentration is changed, the
pump automatically recalculates the infusion rate
based on the new concentration value.
6060™ Multi–Therapy Pump Operator's Manual
prompt, enter the code:
6
0
RE-PROGRAM!
5
because
5-9