Reviewing Profile Parameters; Powering The Pump Off - baxter 6060 2M9832 Operator's Manual

Multi–therapy ambulatory infusion pump
Table of Contents

Advertisement

Basic Operation

Reviewing Profile Parameters

Note:

Powering the Pump Off

07-19-12-353
To review the current programmed parameter values:
If the pump is already running, press RUN/HOLD to
1.
temporarily stop the infusion.
Press NO/REVIEW . The pump displays:
2.
REVIEW PROFILE
PARAMETERS?
Press YES .
3.
The pump displays the current option and limit
settings, then the programmed infusion parameters.
The pump displays change automatically, but you
may advance through them more quickly by pressing
YES as each value is displayed.
When all parameters have been displayed, the pump
displays the Final Confirmation or
screen.
Press RUN/HOLD to start or continue the infusion.
4.
To power off when an infusion is complete,
press
To power off before an infusion is complete,
press
Information for the incomplete program is stored in
memory. The next time the pump is turned on, it will
prompt you to resume the previous infusion.
6060™ Multi–Therapy Pump Operator's Manual
ON/OFF PUMP
.
HOLD
, then press
***HOLD***
ON/OFF PUMP
.
3-17
3