Setting Volume Accrual - baxter 6060 2M9832 Operator's Manual

Multi–therapy ambulatory infusion pump
Table of Contents

Advertisement

Delivery Options and Advanced Functions

Setting Volume Accrual

07-19-12-353
Volume Accrual allows total volume infused to
accrue over repeated infusions in the Intermittent and
PCA profiles. When enabled, it lets you monitor total
volume/medication infused over the duration of a
therapy. The Volume Accrual function is accessible
from the profile-specific programming screen.
If Volume Accrual is not enabled, the volume infused
is reset to zero with each infusion. To select Volume
Accrual:
Turn the pump on.
1.
When the pump displays:
2.
YES to Program
RUN to Repeat
enter the code:
6 4
4
The pump displays:
SELECT ALLOWED
PROFILES?
Press NO . The pump displays:
3.
ENABLE VOLUME
ACCRUAL?
Press YES to enable volume accrual or NO to
4.
disable the feature. The pump displays:
YES to Program
RUN to Repeat
5.
Continue with programming or turn the pump off.
Your settings remain in effect until they are reset.
6060™ Multi–Therapy Pump Operator's Manual