Prior to using this pump, read this manual carefully to fully understand the pump’s functionality and to ensure safe and proper operation. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
Page 2
There are risks associated with using anything other than the recommended sets with this device. Sets designated for use with this device are identified in Chapter 3, Administration Sets. Baxter’s warranty on this device will be null and void and Baxter will assume no responsibility for incidents which may occur if the product is not used in accordance with product labelling.
Programming Screens ................2-10 Main Display Icons .................. 2-11 Label Location ....................2-12 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 4
Panel Lockout ..................... 3-21 Panel Lockout Overview ................. 3-21 Using Panel Lockout................3-21 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Page 5
Using the Optional Prime Function ..............5-17 Overview ....................5-17 Priming the Administration Set..............5-17 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 6
Unloading the Administration Set after Channel Failure ......6-12 Resetting Manual Tube Release after Channel Failure ......6-12 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Page 7
..........Index-1 Printed from the Baxter Intranet.
■ Thermoplastic Synthetic Rubber ■ Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 1. Pump housing components may be made of PC/ABS or Acetal. Some pumps may contain PC/ABS and Acetal components.
Deerfield, IL 60015 USA Made in Singapore Authorised Baxter S.A. Representative: B-7860 Lessines, Belgium Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 10
Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
Long Term Care ■ •Cardiac Catheter Lab Nursing Homes ■ •Emergency Room •Burn Unit/Trauma •Oncology Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 12
See “Technical Specifications,” 8-1 for configurable features and default settings. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 13
Introduction Overview U.S.: Uses Baxter standard administration sets equipped with keyed On/Off Clamps. EEC COUNTRIES: Uses Baxter Green Label C96XX administration sets equipped with keyed On/Off Clamps. See “Recommended Administration Sets,” 3-4. Automatic Tube Loading Variety of Source Containers Micro and Macro Rate Range 0.1 mL/hr to...
Page 14
Configurable Pump Personality Feature Sets Battery Charge Level Indicator Nurse Call Communications Port Diagnostics Functions Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
When disposing of this device or the administration sets designed ■ for use with the device, adhere to local regulations and guidelines. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764...
Page 16
CAN/CSA C22.2 NO. 601.1. 5R78 Labeling abbreviations: ■ COMM. PORT Communications Port VOL. Volume CONT. Contrast Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
This pump should be used only with Baxter accessories specified in “Using the ! WARNING ! Optional Accessories,” 5-22. There are risks associated with using anything other than the recommended accessories with this pump.
Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 19
Introduction Safety Summary When infusing through a central line catheter, Baxter recommends using sets Caution with a Luer lock adapter. Follow the cleaning schedule and methods defined under “Cleaning Overview,” Caution 7-1 to ensure proper maintenance of the device. Do not clean, disinfect, or sterilize any part of the pump by autoclaving or with Caution ethylene oxide gas.
Certain combinations of error codes can result in default to the English language. If this should occur, dis- continue using the pump and contact your authorized Baxter Ser- vice Center for assistance. Note: U.S.
Chapter 2 Overview This chapter describes the operator controls and indicators on the pump. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Main Body Main Body MAIN BODY PUMP MODULE Figure 2-1 Front View Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Page 23
This key powers on the pump. If the pump is on, pressing this key powers down all parts of the pump except memory and circuits that monitor information such as battery charge level. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764...
Page 24
The pump can be monitored by a computer connected to the COMM port on the rear. COMPUTER CONTROL FOR FUTURE USE. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Pop-Up Window This message box is used to provide additional information that does not require user response. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Displays the activity of the pump module during specific operations such as tube loading. Displays alarm or alert conditions when they occur. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 27
This red LED flashes on and off during an alarm condition and remains on continuously during a failure condition. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764...
Communication port An RS232/RS423 interface enables optional nurse call and communication functions. The pump can be monitored by a computer. Refer to the Colleague Pump Computer Monitoring Guide, available from your Baxter customer representative, for information on computer monitoring. Fuse holders The pump’s fuses are located inside.
Feature Set Operating Programming State Icon Values Label Prompt Figure 2-5 Main Display Information Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Select soft key. Pop-up Window Figure 2-7 Programming Modes Menu Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 2-10 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Use the Page Up and Page Down soft keys to page through the list. MORE ARROW Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 2-11 07-19-D4-764...
Label Location The pump’s labels provide additional information about the pump. Figure 2-8 shows the location of the pump’s labels. If any labels are missing or damaged, contact your local Baxter Service Center for replacement information. WORDMARK MAINTAIN BATTERY CHARGE (OPTIONAL)
The mounting clamp can be attached to an IV pole, headboard or footboard. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764...
Turn the mounting clamp knob clockwise to close the clamp arm. Figure 3-2 Headboard (horizontal) Tighten until secure. Mount Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Continue with Steps 2 and 3 of the appropriate mounting procedure. Figure 3-3 Changing the Mounting Bracket Orientation Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Clamps that have an "s" after the code number, or are labeled as pump compatible. If you have questions about administration set compatibility, contact the Baxter Product Information Center at the number shown on the administration set labeling. There are risks associated with using anything other than the recommended administration sets with this device.
Prepare the primary infusion container following the manufacturer’s directions for use. Attach an appropriate Baxter administration set to the solution container and prime the administration set following the directions for use. Ensure all air is expelled from the administration set.
Figure 3-6 Pump Module Display Audible speaker is not heard during self-diagnostic test. Figure 3-7 LEDs Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 39
Feature Sets,” 5-23 for additional information. During Self-Test Figure 3-9 Power On Screen After Self-Test Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
This alarm occurs if the ON/OFF CHARGE key is pressed before the pump finishes the loading or unloading process. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
Ensure that all steps have been properly performed. If flow is still observed, remove the pump from service and contact a Baxter authorized service person. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
If the loading mechanism is disabled (for example, the battery has depleted), see “Using the Manual Tube Release,” 3-11 for instruc- tions on unloading the administration set manually. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-10 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Figure 3-13 Using the Manual Tube Release Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-11 07-19-D4-764...
You cannot use the pump until the Manual Tube Release is reset and the pump is powered off and back on. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-12 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Program the desired flow rate (in mL/hr) using the keypad (Figure 3-16). Figure 3-16 Rate-Volume Programming Screen Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-13 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 46
Programming values for each field are also stored when the pump is in stop mode and then powered off. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-14 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Press the key. START Figure 3-20 Infusion Stopped Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-15 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
KVO alert mode after the piggyback infusion has been completed. Figure 3-21 Select Callback Yes or No Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-16 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
KVO rate if no primary infusion exists. Figure 3-23 Enter Piggyback VTBI Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-17 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
To restart the piggyback infusion, press the Piggyback soft key followed by the START key. Figure 3-25 Press START Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-18 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Primary infusion programs cannot be altered when a piggyback infusion is running. Press START to accept new Rate Figure 3-27 New Rate Running Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-19 07-19-D4-764...
Last Volume Cleared displayed in the field. Figure 3-28 Volume History Screen Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-20 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
If disabled keys are pressed when the keypad is locked, the Panel Locked Pop-up is displayed (Figure 3-30). Figure 3-30 Front Panel Locked Invalid Key Pressed Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-21 07-19-D4-764...
Page 54
Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
Options accesses a menu as shown in Figure 4-2. Figure 4-1 Main Display Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Figure 4-3 Options Menu Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
See “Configurable Options,” 9-1 for factory settings. Figure 4-5 Highlight Name of Current Personality Feature Set Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Figure 4-7 Downstream Occlusion Values Pop-Up Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Use the ר Select soft key to access it. Figure 4-8 Options Menu Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
(Figure 4-10). See the Colleague Global Service Manual for details. Figure 4-10 Options Menu Select Configuration/Service Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Additional programming functions and utilities ■ Using optional accessories ■ Pump Personality Feature Sets ■ Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
See “Selecting a Label,” 5-3 for selecting labels for display. Figure 5-2 Main Display Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
To clear a label, use the same procedure, but select from the label list. Figure 5-4 Label List Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Drug Amount and Diluent Volume fields are cleared. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
(Figure 5-5) to display the available modes (Figure 5-6). Figure 5-5 Primary Programming Screen Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 66
To clear the parameters, press the Clear Settings soft key (Figure 5-7). Figure 5-7 Clear Old Settings Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Page 67
Diluent Volume (Figure 5-9) and use the Ø field. Figure 5-9 Enter Drug Amount Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 68
The pump will then calculate the Dose. Figure 5-11 Value “HIGH” Relieved Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Page 69
See “Label Library,” 5-2 for details. Figure 5-13 Enter Volume to be Infused Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 70
(Figure 5-15). Figure 5-15 Confirm Settings Prompt Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-10 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
The Programming Modes screen is displayed (Figure 5-18). Figure 5-17 Primary Programming Screen Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-11 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 72
Program each field as described in “Programming a Dose Independent of Patient Parameters,” 5-5. Figure 5-19 Clear Stored Settings Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-12 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 73
Press the START key to begin the infusion. Figure 5-21 Confirm Settings Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-13 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
To use other programming functions, press the Change Mode soft key (Figure 5-23). Figure 5-23 Restore Settings Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-14 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Volume-Time or Piggyback Volume-Time , then press the Select soft key. Figure 5-24 Change Mode Screen Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-15 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 76
Time Remaining instead of the time duration you entered. Figure 5-26 Enter Time Duration Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-16 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Programming screen. From the Programming screen, press the Change Mode soft key. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-17 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 78
Prime soft key. Figure 5-29 Ready to Prime Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-18 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
If the pump is powered off, program information is retained for 5 ■ hours. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-19 07-19-D4-764...
YES shown on the pop-up to place the × pump into Standby mode. Figure 5-32 Standby Pop-Up Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-20 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Main Display. The pump exits Standby and reverts to the programming mode in effect when the pump was placed into Standby. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-21 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Using the Optional Accessories Using the Optional Accessories This pump should be used only with Baxter accessories specified in “Using the ! WARNING ! Optional Accessories,” 5-22. There are risks associated with using anything other than the recommended accessories with this pump.
DL Event History Download Kit (product code 2M8317, U.S. only). Personality Feature Sets Overview To meet our customers’ needs, Baxter designed the Colleague Pump to be highly adaptable to each care setting. Throughout this manual, references are made to specific programming functions described as configurable options.
Figure 5-35 Configuration Screen Example Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-24 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
To silence the alert tone for two minutes, press the key. Alarm Silence Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Indication: The pump is in Advance Air mode. The Advancing Air screen is displayed (Figure 6-2). Figure 6-2 Advancing Air Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Page 87
Figure 6-4 Advance Air Volume Limit Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
AIR/label Indication: The pump has detected an air bubble based on a configured setting. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 90
When the pump detects fluid, a fluid detected icon is displayed (Figure 6-7). Figure 6-6 Advancing Air Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 91
! WARNING ! always close the regulating clamp on the administration set before removing the administration set from the pump. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764...
Check the current Personality feature set configuration. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 93
• If alarm occurs again, a hardware problem may exist. Take the pump out of • A hardware problem may have service and have it inspected by a Baxter occurred. authorized service person. Tube Not Loaded • Press the Open key to reset the alarm.
Main Display if necessary by pressing the Primary Piggyback soft key. Figure 6-9 Channel Failure Status Line Display Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 6-10 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Page 95
Alarm LED will light, and the audible tone will occur. Figure 6-11 Channel Failure Display Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 6-11 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Manual Tube Release is reset and the pump is powered off and back on. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 6-12 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
The pump’s design safeguards against fluid spillage into the pump module. Contact your Baxter Service Center for assistance immediately if fluid enters the tubing channel. The tubing channel should be cleaned as soon as possible by authorized service personnel to minimize potential difficulties caused by fluid pooling and drying on the mechanism.
If the pump cannot be cleaned using the basic methods described earlier or components are missing or damaged, discontinue use and notify the appropriate authorized service personnel. To contact Baxter for authorized service or repair, see “Service Information,” 10-2. Check Action Perform as required but recommended after every use.
Service Center (see “Warranty and Service Information,” 10-1). Battery Disposal The pump contains sealed lead acid batteries which should not be disposed of in the trash. They may be returned to your authorized Baxter Service Center marked “FOR DISPOSAL.” Printed from the Baxter Intranet.
34 days (batteries removed from pump) 1. Baxter recommends removing the batteries if the pump is to be stored for more than 90 days. Storage of the pump at temperatures greater than 24°C (75°F) will hasten battery discharge which may shorten the batteries’ overall operating life.
U.S.: 2M8151 U.K.: 2M8151K Single Channel Shuttle Volumetric Infusion Pump Administration Sets U.S.: Standard Baxter administration sets equipped with keyed On/Off Clamps Latin America: Baxter “Green Label” C96XX administration sets equipped with keyed On/Off Clamps or Standard Baxter administration sets equipped with keyed On/Off Clamps EEC Countries: Baxter “Green Label”...
Page 102
465 mmHg (9 psig) 620 mmHg (12 psig) 775 mmHg (15 psig) Maximum COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
Consult the infusion system or accessory manufacturer’s instructions for use before proceeding. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-E4-764...
Figure 8-1 Startup Graph measured is lengthened, all pumps show considerable improvement in Example flow accuracy. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Lines are then drawn to connect the plotted points to create the ■ trumpet curve. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Average of 5 pumps Figure 8-4 Trumpet Graph, 2nd Hour of Delivery, 1 mL/hr Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
37.5 12.5 Figure 8-7 Delivery Startup, First Two Hours, 25 mL/hr Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 108
Average of 5 pumps Figure 8-10 Trumpet Graph, 72nd Hour of Delivery, 25 mL/hr Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 109
The information in the following tables represents laboratory testing conducted per Sub-Clause 51.6b of IEC 60601-2-24 Part 2. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764...
Page 110
For flow rate < 1.0 mL/hr, worst case volume is 25% of set VTBI or 1.33 mL, whichever is greater. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
When enabled, the following items affect the entire pump: Personality Names and Configuration Settings ■ Set Time and Date ■ 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
Custom labels can also be programmed if desired. (See “Label Library,” 9-5.) Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 113
• Label (The Label Library feature and individual labels must be enabled). • Volume Infused (mL) Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 114
Number of Auto Restarts 0 to 9 after a Downstream Occlusion Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Analgesics / Anesthetics Alfentanil HCL ALFENTAN Atracurium Besylate ATRACURI Doxapram HCL DOXAPRAM Droperidol DROPERID Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 116
Doxorubicin HCL DOXORUBI Etoposide ETOPOSID Fluorouracil FLUOROUR Ifosfamide IFOSFAMI Methotrexate Sodium METHOTRE Plicamycin PLICAMYC Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Page 117
LABETALO Metaraminol Bitartrate METARAMI Milrinone Lactate MILRINON Norepinephrine Bitartrate NOREPINE Phenylephrine HCL PHENYLEP Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Custom labels appear in the list in alphabetical order. See the Colleague Global Service Manual for instructions on configuring custom labels. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
There are risks associated with using anything other than the recommended Baxter designated IV administration sets with this device. Baxter’s warranty to repair or replace the product will be null and void if this product is used contrary to the directions for use contained in the labeling or if used with non-recommended sets.
In the event that your pump needs to be returned for service, obtain a Return Authorization by calling 1-800-THE-PUMP or your local service center. Shipping costs for all pumps returned to Baxter shall be paid for by the customer. The pump must be packed in its original container or in another Baxter approved container that will provide adequate protection during shipment.
Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. Index-1 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 122
5-11 headboard mounting of pump entering diluent volume history 3-20 entering dose Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. Index-2 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Page 123
IV pole mounting soft location start overview Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. Index-3 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 124
3-21 prime active alarm 5-18 prime warning procedure 5-18 3-12 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. Index-4 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Page 125
6-10 specifications incomplete program alarm environmental lithium battery low alert Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. Index-5 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
Page 126
5-22 warranty 10-1 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. Index-6 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-D4-764...
Page 127
Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.