Giving a Clinician-Activated Dose

4-50
If the patient requests a bolus but the mandatory
interval between bolus doses has not elapsed or the
number of doses allowed per hour have already been
given, the requested bolus dose will not be given.
The pump records the request in the PCA data log,
displays
BOLUS DOSE REQUESTED
times just as if a dose is being given.
You can deliver a clinician-activated bolus dose at
any time during a PCA infusion, and at any level of
security. The function is accessible via a numeric
code.
After a clinician-activated dose is delivered, the
cumulative volume infused reflects all volume
infused, including the clinician-activated dose
volume.
During a PCA infusion, press RUN/HOLD to pause
1.
the infusion.
The pump displays the current basal rate and
cumulative volume infused, alternating with:
***
To access the Clinician Dose function, enter the code:
2.
5
1
The pump displays:
Dose:
Enter the desired clinician bolus dose and press YES
3.
to accept the entered dosage.
6060™ Multi–Therapy Pump Operator's Manual
HOLD
***
5
1
CLINICIAN
0.0 mg
and beeps four
07-19-12-353