Programming PCA Infusions
Note:
Note:

Set PCA Profile Titration Limits

4-42
The pump displays:
5.
Medication Vol
Limit: 0.0 mg
Press YES to accept the displayed medication
6.
volume limit, or enter a new volume limit and press
YES to accept.
7.
Proceed to "Set PCA Profile Titration Limits," 4-42.
The Medication Volume limit should account for all
hours of basal rate plus the bolus doses allowed dur-
ing the specified time limit. The pump calculates the
medication volume limit by including the sum of the
basal volume plus the delivered bolus volumes, but
does NOT include the volume of any Clinician-Acti-
vated boluses or loading dose.
If you entered a basal rate of zero and answered NO
to
DEMAND BOLUS DOSE?
ALERT Program
Basal &/or Bolus
Press YES or NO to return to step 1 of this proce-
dure. You MUST either program a continuous basal
rate or allow bolus doses to be administered.
To allow titration during a PCA infusion, press YES
1.
at the prompt and continue on to set parameters.
SET TITRATION
If you do not wish to allow titration, press NO and go
2.
to "Program Optional Loading Dose," 4-44.
6060™ Multi–Therapy Pump Operator's Manual
, the pump displays:
LIMITS?