Program Optional Loading Dose; Check Programmed Parameters - baxter 6060 2M9832 Operator's Manual

Multi–therapy ambulatory infusion pump
Table of Contents

Advertisement

Programming PCA Infusions

Program Optional Loading Dose

Note:

Check Programmed Parameters

4-44
MAXIMUM Bolus
Doses/hr:
Press YES to accept the displayed maximum number
8.
of bolus doses per hour allowed for titration, or enter
a new value and press YES to accept.
A loading dose can be given as the pump begins
infusing. If you press NO to administering the loading
dose, it cannot be given later.
The pump displays:
Loading Dose?
If you do not want to program a loading dose, press
9.
NO . To program a loading dose, press YES. The
pump displays:
Dose:
Press YES to accept the displayed loading dose, or
10.
enter a new value and press YES to accept.
To allow you to check or change the values you
programmed, the pump displays:
CHECK or CHANGE
PCA VALUES?
alternating with:
6060™ Multi–Therapy Pump Operator's Manual
X
PROGRAM
Loading
XXX mg
07-19-12-353

Advertisement

Table of Contents
loading

Table of Contents