INSTRUCTIONS FOR USE
b

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EasySpray and is the answer not in the manual?

Questions and answers

Summary of Contents for baxter EasySpray

  • Page 1 PRAY P R E S S U R E R E G U L A T O R INSTRUCTIONS FOR USE...
  • Page 2 PRAY P R E S S U R E R E G U L A T O R Sterile filter luer-lock connectors sensing line pressure line connector connector Pressure gauge IV-pole clamp Gas supply line connector with filter On/Off switch battery compartment Pressure Low-battery...
  • Page 3: Table Of Contents

    INSTRUCTIONS FOR USE TABLE OF CONTENTS 1. Description 2. Warnings and Precautions 3. Storage 4. Operation 5. Cleaning / Maintenance 6. Terms of Guarantee 7. Return of Devices 8. Waste Disposal of Product 9. Technical Data 10. Manufacturer and Distributor 11.
  • Page 4: Description

    1. DESCRIPTION The EASYSPRAY Pressure Regulator controls and releases pressurized gas provided by a propellant gas source. The EASYSPRAY device is to be used only with spray sets specified for use with this pressure regulator. Only qualified operating room personnel should operate the device, having first read this manual carefully.
  • Page 5: Warnings And Precautions

    • DO NOT throw or drop the device! • Do not use the device in the case of visible damage caused by improper handling. Return damaged devices to Baxter for repair. 3. STORAGE • Storage and Transport Conditions: Temperature -10º to 45ºC, 60% relative air humidity.
  • Page 6 Baxter in accordance with Section 7 (RETURN OF DEVICES). 4.3 OPERATION • The clamp on the back side of the EASYSPRAY device allows the device to be mounted onto an IV-pole or a bed-rail. Firmly set screws to assure stable connection to either the IV-pole or bed-rail.
  • Page 7: Cleaning / Maintenance

    5. CLEANING / MAINTENANCE 5.1 CLEANING The EASYSPRAY housing can be cleaned with standard operating room disinfectants. In order for the device to operate safely and in accordance with the intended use, the device must be cleaned following use and maintained free of dirt and other debris.
  • Page 8: Terms Of Guarantee

    6. Remove the retaining washer. 7. Mount the pressure control knob. 8. Close/Cover the sensing line connector to start gas flow through the EASYSPRAY device and set the output pressure to 2.8 bar. 9. Check the function of the EASYSPRAY device by starting it repeatedly, i.e., covering and uncovering the sensing line connector.
  • Page 9: Return Of Devices

    Reasons for return: 1. Results of the first-time use tests are unsatisfactory/unsafe 2. Defective device Please contact Baxter customer service to obtain a RETURN AUTHORIZATION (RA) NUMBER prior to returning. Product returns will not be accepted without a Return Authorization number.
  • Page 10: Electromagnetic Compatibility Information

    WARNING: When servicing the instrument, use only replacement components, cables and accessories authorized by Baxter and be sure to replace all shields, covers, screws, and gaskets in their exact locations as described in the service manual. Failure to do so may result in increased emissions or decreased immunity of the instrument.
  • Page 11 To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the EasySpray Pressure Regulator is used exceeds the applicable RF compliance level above, the EasySpray Pressure Regulator should be observed to verify normal operation.
  • Page 12 PRAY P R E S S U R E R E G U L A T O R Baxter Healthcare Corporation Westlake Village, CA 91362 1-800-423-2090 US: Rx Only 6299610EH04...