Specifications
Power Meter Supplemental Characteristics
For Agilent E-series power sensors
In FAST mode, within the range -50 dBm to +17 dBm, for a 10 dB decreasing power
step, the settling time is 10 ms
1
When a decreasing power step crosses the power sensor's auto-range switch point, add
25 ms. Refer to the relevant power sensor manual for further information.
For Agilent E-series power sensors in normal and x2 speed modes, manual filter,
10 dB decreasing power step:
Number of Averages
Settling Time (s)
(Normal Mode)
Response Time (s)
(x2 mode)
5-10
Table 5-5: Settling Time
1
2
4
8
0.07
0.12
0.21
0.4
0.04
0.07
0.12
0.21
1.
16
32
64
1
1.8
3.3
0.4
1
1.8
128
256
512
1024
6.5
13
27
3.4
6.8
14.2