Power Meter Operation
Making Measurements with the Agilent E-Series Power Sensors
Example
To make a measurement using an Agilent E-series power sensor. The frequency of the
signal to be measured is 100 MHz.
Disconnect the power sensor from any source.
Zero
Press
.
Cal
Press
.
Zero
Connect the power sensor to the POWER REF output.
Press
.
Cal
Frequency
Press
Cal Fac
to enter 100. Press
Connect the power sensor to the signal to be measured.
The measurement result is now displayed.