Agilent Technologies E4418B User Manual page 91

Power meter
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

To select the format of the display in the currently selected window:
Meas
1.
Press
,
Display Format
Setup
2.
Press
Meter Dgtl Anlg
Pressing this key toggles between a digital and analog display.
To select the range of the analog display in the currently selected window:
Meas
1.
Press
,
Display Format Anlg Mtr Scaling
Setup
2.
The current setting of the maximum and minimum values displayed on the
analog meter are displayed under the
3.
To change either of these settings press the appropriate softkey. The power
meter displays the current value in a pop up window. Modify this value (see
below) until the desired value is displayed.
Use
n
positioned.
Use
n
4.
The choices available depend on the unit of measurement selected and the
current measurement mode. If the measurement unit selected is logarithmic,
then the choice will be
linear, then choose the appropriate softkeys from the following menus:
More
Pressing
Increment Multiplier
increases or decreases the multiplier shown in front of
or
after the correct multiplier has been selected confirms the entry.
W
To select both an analog and digital display press
displayed. Select the format of one window to analog and the other to digital as
previously described.
Agilent E4418B User's Guide
.
to highlight the type of format you require.
,
Max
or
to modify the digit on which the cursor is currently
or
to move to other digits.
or
. If the measurement unit selected is
dB
dBm
%
Cancel
%
Increment
Multiplier
Decrement
Multiplier
Cancel
Power Meter Operation
Selecting a Digital or Analog Display
.
and
softkeys respectively.
Min
mW
µW
nW
Cancel
W
Increment
Multiplier
More
Decrement
Multiplier
Cancel
or
Decrement Multiplier
or
. Pressing
%
W
until two windows are
%
2-49