Power Meter Operation
Making Measurements using Sensor Calibration Tables
The following lists detail the contents of the predefined sensor calibration tables.
DEFAULT
RCF
100
0.1 MHz
100
110 GHz
100
Agilent 8481A
RCF
100
50 MHz
100
100 MHz
99.8
2 GHz
99
3 GHz
98.6
4 GHz
98
5 GHz
97.7
6 GHz
97.4
7 GHz
97.1
8 GHz
96.6
9 GHz
96.2
10 GHz
95.4
11 GHz
94.9
12.4 GHz
94.3
13 GHz
94.3
14 GHz
93.2
15 GHz
93
16 GHz
93
17 GHz
92.7
18 GHz
91.8
2-24
Agilent 8482A
RCF
98
0.1 MHz
98
0.3 MHz
99.5
1 MHz
99.3
3 MHz
98.5
10 MHz
98.5
30 MHz
98.1
100 MHz
97.6
300 MHz
97.5
1 GHz
97
2 GHz
95
3 GHz
93
4.2 GHz
91
Agilent 8483A
RCF
94.6
0.1 MHz
94
0.3 MHz
97.9
1 MHz
98.4
3 MHz
98.4
10 MHz
99.3
30 MHz
98.7
100 MHz
97.8
300 MHz
97.5
1 GHz
97.2
2 GHz
96.4