Rx Pacing
o
Press this softkey to toggle between enabling and disabling the
Xon/Xoff receiver software handshake. When enabled "Xon" is
highlighted, otherwise "None" is highlighted.
RTS/CTS
o
Press this softkey to obtain a pop up window which allows you to
select one of the following:
OFF - Sets the RTS signal line permanently low.
ON - Sets the RTS signal line permanently high.
IBFull - Sets the RTS signal line high while the receiver
DTR/DSR
o
Press this softkey to obtain a pop up window which allows you to
select one of the following:
OFF - Sets the DTR signal line permanently low.
ON - Sets the DTR signal line permanently high.
IBFull - Sets the DTR signal line high while the receiver
This key is greyed out if RS422 interface is selected.
Echo
v
Press this softkey to toggle Echo "On" or "Off". When Echo is
enabled, received characters are transmitted back to the sender.
Tables
Press this softkey to access a menu which allows you to select
between sensor calibration tables and frequency dependent offset
tables.
Sensor Cal Tables
n
Press this softkey to access a menu and screen which allows you to
select and edit sensor calibration tables. Use the
hardkeys to scroll through the displayed sensor calibration tables.
Sensor calibration tables are only required for the Agilent 8480 series
power sensors.
Edit Table
u
Agilent E4418B User's Guide
Front Panel Menu Reference "System/Inputs" Menu
buffer can accept more data and sets RTS low when
the data buffer is full. The transmitter is inhibited
when CTS is low.
buffer can accept more data and sets DTR low when
the data buffer is full. The transmitter is inhibited
when DSR is low.