-310
System error;Sensor EEPROM Read Failed - unknown EEPROM table
format
This indicates a failure with your Agilent E-series power sensor. Refer to your
power sensor manual for details on returning it for repair.
-310
System error;Sensor EEPROM < > data not found or unreadable
Where < > refers to the sensor data block covered, for example, Linearity, Temp -
Comp (temperature compensation).
This indicates a failure with your Agilent E-series power sensor. Refer to your
power sensor manual for details on returning it for repair.
-310
System error;Option 001 Battery charger fault
The power meter is connected to an ac power source, the battery is not fully
charged and it is not charging.
-310
System error;Sensors connected to both front and rear inputs.
You cannot connect two power sensors to the one channel input. In this instance
the power meter detects power sensors connected to both it's front and rear
channel inputs.
-321
Out of memory
The power meter required more memory than was available to run an internal
operation.
-330
Self-test Failed;
The -330, "Self-test Failed" errors indicate that you have a problem with your
power meter. Refer to "Contacting Agilent Technologies", on page 2-72 for details
of what to do with your faulty power meter.
-330
Self-test Failed;Measurement Channel Fault
Refer to "Measurement Assembly", on page 2-69 if you require a description of
the Measurement Assembly test.
-330
Self-test Failed;Option 001 Battery requires replacement
The option 001 battery is not charging to a satisfactory level and should be
replaced.
-330
Self-test Failed;RAM Battery Fault
Refer to "RAM Battery", on page 2-69 if you require a description of the battery
test.
Agilent E4418B User's Guide
Error Messages
Error Messages
4-9