Error Messages
Error Messages
-158
String data not allowed
A character string was received but is not allowed for the command. Check the list
of parameters to verify that you have used a valid parameter type.
For example, LIM:STAT 'ON'.
-161
Invalid block data
A block data element was expected but was invalid for some reason. For example,
*DDT #15FET. The 5 in the string indicates that 5 characters should follow,
whereas in this example there are only 3.
-168
Block data not allowed
A legal block data element was encountered but not allowed by the power meter at
this point.
For example SYST:LANG #15FETC?.
-178
Expression data not allowed
A legal expression data was encountered but not allowed by the power meter at
this point.
For example SYST:LANG (5+2).
-211
Trigger ignored
Indicates that <GET> or *TRG, or TRIG:IMM was received and recognized by the
device but was ignored because the power meter was not in the wait for trigger
state.
-213
Init ignored
Indicates that a request for a measurement initiation was ignored as the power
meter was already initiated.
For example, INIT:CONT ON
-214
Trigger deadlock
TRIG:SOUR was set to HOLD or BUS and a READ? or MEASure? was
attempted, expecting TRIG:SOUR to be set to IMMediate.
-220
Parameter error;Frequency list must be in ascending order.
Indicates that the frequencies entered using the MEMory:TABLe:FREQuency
command are not in ascending order.
4-6
INIT.
Agilent E4418B User's Guide