The Display Layout

The following figure details the display layout when two measurement windows are
displayed, one analog and one digital. However it is possible using the
display just one measurement window.
14
1
2
3
4
5
1.
The status reporting line displays five fields, three associated with either
GP-IB, RS232 or RS422 status and two associated with error and warning
conditions. The first field displays either "RMT" (remote, GP-IB, RS232 or
RS422 operation) or "LCL" (local, front panel operation).
For GP-IB operation, the second field displays "TLK" if the power meter is
addressed to talk or "LSN" if it is addressed to listen. The third field
indicates an "SRQ" (service request).
For RS232 and RS422 operation, the second field displays "RX" when data
is being received. The third field displays "TX" when the power meter is
transmitting data.
The fourth field indicates "ERR" for any error conditions. The last field is
used to report error and warning messages.
2.
The measurement data is displayed in either one or two rectangular windows
depending on the setting of
between a one or two window display. When two windows are displayed and
this hardkey is pressed the single window then displayed is the one which
was previously highlighted with the shadowed box. On the two window
Agilent E4418B User's Guide
13
12
11
10
9
. Pressing
allows you to toggle
Getting Started
The Display Layout
key to
8
7
6
1-7