4.1
REFERENCE CONDITIONS
Quantities of Influence
Temperature:
Relative humidity:
Supply voltage:
Frequency range of the applied signal:
Sine wave:
Peak factor of the applied alternating signal:
Position of the conductor in the clamp:
Adjacent conductors:
Alternating magnetic field:
Electric field:
4.2
SPECIFICATIONS UNDER THE REFERENCE CONDITIONS
Accuracy is expressed in ± (x% of the reading (R) + y counts (ct)).
4.2.1
DC Voltage Measurement
Measurement range
Specified
measurement range
Accuracy
Resolution
Input impedance
Note (1)
The display indicates "+OL" above + 2000V and "-OL"
below – 2000V, in REL mode. The "-" and "+" signs are managed.
Above 1000V, a repetitive beep indicates that the voltage being
measured is greater than the safety voltage for which the device is
guaranteed. The display indicates "OL."
4 SPECIFICATIONS
0.00 to 59.99V
0 to 100% of the measurement range
from 0.00V to 5.99V
±(1% R + 10cts)
from 6.00V to 59.99V
±(1% R +3cts)
0.01V
31
Reference Conditions
23°C ±2°C
45% to 75%
9.0V ±0.5V
45 to 65Hz
pure
√2
centered
none
none
none
60.0 to 599.9V
±(1% R +3cts)
0.1V
10MΩ
(1)
600 to 1000V
1V