6.8 ELECTRICAL SAFETY

The devices comply with IEC/EN 61010-2-030 or BS EN 61010-2-030 for a
voltage of 600 V measurement CAT III, pollution degree 2.
The devices comply with BS EN 62749 for EMF. Product is intended for
professional use only.
Battery load between terminals V1 and N: 600 V surge CAT III, pollution degree 2.
The test leads and alligator clips comply with standard IEC/EN 61010-031 or
BS EN 61010-031.

6.9 ELECTROMAGNETIC COMPATIBILITY

Emissions and immunity in industrial environment compatible with IEC/EN
61326-1 or BS EN 61326-1.
With the AmpFlex
0.5 % of full scale with a maximum of 5 A.

6.10 RADIO EMISSION

The devices comply with the RED Directive 2014/53/EU and FCC regulations.
FCC certification number for the Wi-Fi: FCC QOQWF121

6.11 MEMORY CARD

Transferring a large amount of data from the SD card to a PC can take a long
time. In addition, some computers can have difficulty processing such large
amounts of information, and spreadsheets only accept a limited amount of data.
To transfer data faster, use the SD/USB card adapter.
Do not exceed 32 recordings on the SD card or the SD card may become full.
The maximum size of a recording is 4 GB. The recording duration is unlimited
(> 100 years).