Using the M1970 Series Mixers with the N9030A, N9020A, and
N9010A Signal Analyzers (Option EXM)
Refer to
configuration requirements.
The following examples explain how to connect the external mixers to the signal
analyzer and how to use the signal-identification functions.
For additional information regarding use with a particular X-Series analyzer, refer
to that analyzer's User's Guide and Programmer's Reference Guide.

Equipment setup

Table 2-1
Equipment setup for M1970 Series mixers and N9030A, N9020A, and N9010A
signal analyzers (Option EXM)
Step
Action
a Connect an SMA cable from
the mixer to the spectrum
SMA connection
analyzer EXT MIXER front panel
SMA.
a Connect a USB cable from the
USB connection
mixer to the spectrum analyzer.
26
"Spectrum Analyzer Retrofit Requirements"
Notes
The torque for the SMA cables and adapters should not exceed 8 in-lbs.
The LO/IF connection between the mixer and the signal analyzer must be
in place before the USB cable from the mixer is connected. Connecting
the USB cable automatically triggers the LO adjustment, and if the LO/IF
cable is not connected the adjustment will not complete and an error will
occur.
When a connection is made, the green LED on the mixer lights up
indicating that the mixer has power and the processor inside the mixer is
running. The spectrum analyzer is automatically switched to external
mixing mode, and the mixer model number and mixer option are used to
set the start and stop frequencies for the mixing band.
on page 20 for the
Keysight M1970 Series User's Guide