Sample Gas

This list allows you to specify the Sample gas you want to analyze.
Figure 3-6.4.
Sample Gas menu.
Select Gas
Custom gas
3-8 | 3050-AMS Moisture Analyzer
Config
Sample Gas
Gas List
Press Enter to move to the listing of Sample gases. Scroll up/down the list
to select the correct Sample gas.
If you select the Custom Gas selection and press Enter, you will be
prompted to enter a Gas K Factor. The Gas K Factor is a parameter used
by the flow meter in the 3050-AMS to adjust for the molar heat capacities
of different gases. Nitrogen is used as the Reference gas for the K factors,
and its value is defined as 1.000.
Please contact AMETEK Service for assistance, before running gases that
are not listed in the menu on the 3050-AMS Analyzer.
If a custom gas is entered, the conversion to units of dew point will be
based on an ideal gas model, without any enhancement factor.
Air
Argon
Carbon Dioxide
Custom Gas
Gas K Factor
Freon 12
Freon 114
Freon 116
Helium
Hydrogen
Krypton
Neon
Nitrogen
Oxygen
SF6
Xenon