3050-Am Moisture Analyzer Approvals And Certifications - Ametek 3050-AMS User Manual

Moisture analyzer
Hide thumbs Also See for 3050-AMS:
Table of Contents

Advertisement

Specification
Alarms/Alerts
Environmental Conditions
Utility Requirements
Mounting
Dimensions
Net Weight

3050-AM Moisture Analyzer Approvals and Certifications

Design Classifications
5-2 | 3050-AM Moisture Analyzer
Description
Three (3) independent contact closures of 30 VAC or 60 VDC
maximum, 10 VA or 1 A maximum resistive, for system alarm,
moisture concentration alarm, and data valid. All are fail-safe by
default.
Alarms are available on RS-485 interface.
5 °C to 50 °C (41 °F to 122 °F)
[Optimal performance in low-level applications is obtained when
ambient temperature is maintained at a constant temperature
(±2 °C)]
90 % RH maximum, noncondensing
IEC Pollution Degree 2
Maximum altitude 2,000 meters (6560 feet)
IEC Installation Category II
Indoor use only, IEC 529; IP40
Power Consumption
100–127 VAC, 47–63 Hz, 185 W Maximum
220–240 VAC, 47–63 Hz, 185 W Maximum
OR
Instrument Air
Minimum 483–690 kPa (70–100 PSIG), -40 °C dew point
5-micron filter recommended
19-inch rack
Width:
48 cm (19 inches)
Height:
13 cm (5.25 inches)
Depth:
42 cm (16.5 inches)
9.8 Kg (22 lb)
MET Certified to:
General Safety Requirements:
Complies with:
Class I, Division 2, Groups A,B,C,D T4
EMC Directive:
Industrial Low Voltage Directive:
UL1604 and CSA213
UL3101, CSA1010
UL1604 and CSA213
EN61326-1
EN61010-1
PN 305804902, Rev A