To set up the test equipment and the logic analyzer

Before testing the specifications of the Agilent 1680A,AD-series or 1690A,AD-
series logic analyzer, the test equipment and the logic analyzer must be set up
and configured.
These instructions include detailed steps for initially setting up the required test
equipment and the logic analyzer. Before performing any or all of the tests in this
chapter, the following steps must be done.
Equipment Required
Equipment
Critical Specifications
Pulse Generator
200 Mhz, 2.5 ns pulse width, <600 ps rise time 8133A option 003
Digitizing Oscilloscope ≥ 6 GHz bandwidth, < 58 ps rise time
Digital Multimeter
0.1 mV resolution, 0.005% accuracy
Function Generator
DC offset voltage ±1.5 V

Set up the test equipment

1 Turn on the required test equipment listed in the table above. Let them
warm up for 30 minutes before beginning any test.
2 Set up the pulse generator according to the following table.
Pulse Generator Setup
Timebase
Channel 2
Mode: Int
Mode: Pulse
Divide: Pulse ÷ 2
Period: 5.000 ns
Width: 2.500 ns
Ampl: 0.80 V
Offs: -1.30 V
COMP: Disabled
(LED Off)
3 Set up the oscilloscope:
a Select Setup, then select Default Setup.
Chapter 3: Testing Performance
To set up the test equipment and the logic analyzer
Recommended HP/Agilent
Model/Part
54750A w/ 54751A
3458A
3325B Option 002
Trigger
Channel 1
Divide: Divide ÷ 2
Mode: Square
Ampl: 0.50 V
Delay: 0.0 ps
Offs: 0.00 V
Ampl: 0.80 V
Offs: -1.30 V
COMP: Disabled
(LED Off)
23