Transmitting Data (Only Oxi 340I); Data Transmission Interval (Int 2, Oxi 340I) - wtw Oxi 330i Operating Manual

Handheld meter
Hide thumbs Also See for Oxi 330i:
Table of Contents

Advertisement

Oxi 330i/340i
ba52306e04
03/2008
4.5

Transmitting data (only Oxi 340i)

You have the following possibilities of transmitting data:
One of the following options:
– With the AutoStore function (page 28), measured values are
periodically saved internally (save interval Int 1) and output on the
interface.
– With the Data transmission interval function (Int 2), measured values
are periodically output on the interface (see below).
With the Output data storage function (page 30), calibration data or saved
measured values are output on the interface.
Via the analog recorder output (page 37), measured values are output as
voltage values.
With the KOM pilot communication kit (accessory), data can be transmit-
ted bidirectionally (page 38).
Note
If you connect a recorder (analog output), the output on the digital interface
is switched off.

4.5.1 Data transmission interval (Int 2, Oxi 340i)

The interval to the data transmission (Int 2) determines the chronological in-
terval between automatic data transmissions. After the selected interval ex-
pires, the current data record is transmitted to the interface.
Note
When the AutoStore function is active, the data transmission is performed
according to the setting of the save interval (Int 1). Set the save interval (Int 1)
to OFF to activate the Data transmission interval (Int 2).