B l u e B o x – S t a t u s a n d E r r o r M e s s a g e s
Appendix J – Status and Error Messages
Display
Parameter
No data
Wait
[Parameter]
Calc error
? Sensor
Calc run
<min. Parameter
>max. Parameter
Com error
ADC min
ADC max
Dev error
At the 6-way parameter display (see 8.1.2 Parameter Display 6-way) status and error messages appear short-
ened.
∗
At virtual sensors the underrun and the overrun is marked by a < or >, only the Min. Value and the Max. Value are stored.
At real sensors the underrun and the overrun the measured value is displayed and stored.
GO Systemelektronik GmbH
Description
Sensor sends data.
Sensor sends no data.
The BlueBox is waiting for the first measured value.
Sensor warning, the measured value may be faulty.
AMS formula error
In an AMS formula an unknown sensor is used.
The calculation time in an AMS formula has been exceeded (for or while loop).
Underrun of the minimum measured value as determined in AMS.
Overrun of the maximum measured value as determined in AMS.
Communication error with the sensor
Underrun of the lower limit of the AD converter
Overrun of the upper limit of the AD converter
General device error
Faluner Weg 1
24109 Kiel
www.go-sys.de
Germany
Tel.: +49 431 58080-0
info@go-sys.de
∗
∗
Page 98 / 99
Fax: -58080-11
Need help?